Cleaning and Storage Biosignal devices

Proper procedure for cleaning, maintaining and storing your Biosignal device (Ring, Versatile Bio)

Cleaning and Disinfection

Ring (Running time: 1 minute. Drying time: 2 minutes).

  1. Apply the Clinell wipe to the entire housing of the device. The device should remain wet 
    for 1 minute counting the application time. Do not apply excessive pressure with the 
    wipe, as this may release liquid that could damage the device internally. Do not insert the 
    wipe into the charging hole or sd slot.

  2. Spread the Clinell wipe over the GSR and BVP sensors, always spreading the wipe 
    slightly with the sensors face down to ensure that no liquid leaks inside the device.

Versatile Bio (Run time: 5 minutes. Drying time: 1 minute). 


To disinfect the sensors (GSR, Temperature, ExG...) use a Clinell wipe.

  1.  Apply the wipe to the sensor head. Surfaces must remain wet for 1 minute counting the 
    application time. This application should be done lightly, always spreading the wipe with 
    the sensors upside down to ensure that no liquid leaks inside the device.
  2. Apply the Clinell wipe to the cables, running from the sensor to the connector (do not 
    apply to the inside of the connector, as this could damage the electronics).

To disinfect the effort band sensor:

  1. For the blue fastening strap, immerse in 2% diluted Bomix. Be careful not to immerse the 
    sensor itself, only the fabric band.

To disinfect the amplifier:

  1. Apply the Clinell wipe to the entire housing of the device. Do not apply excessive 
    pressure with the wipe, this may release liquid which could damage the device internally. 
    Do not insert the wipe into the connector, charging hole, sd slot or phot/dig slot.

 

Storage

Keep the Ring on its original packaging when not in use. 

If the device will not be used for an extended period, we recommend discharging the battery to 40-50% capacity.  Storing the device for more than a week at 100% charge could cause permanent capacity loss. 

Recommended battery storage temperature range:
-10ºC to 35ºC (14ºF to 95ºF) for up to 1 month
-10ºC to 30ºC (14ºF to 86ºF) for up to 6 months


Keep the Versatile Bio and every sensor on its original packaging when not in use. 

If the device will not be used for an extended period, we recommend discharging the battery to 40-50% capacity.  Storing the device for more than a week at 100% charge could cause permanent capacity loss. 

Recommended battery storage temperature range:
-10ºC to 35ºC (14ºF to 95ºF) for up to 1 month
-10ºC to 30ºC (14ºF to 86ºF) for up to 6 months

 

Safety data sheets

 