Cleaning and Storage Biosignal devices
Proper procedure for cleaning, maintaining and storing your Biosignal device (Ring, Versatile Bio)
Cleaning and Disinfection
Ring (Running time: 1 minute. Drying time: 2 minutes).
- Apply the Clinell wipe to the entire housing of the device. The device should remain wet
for 1 minute counting the application time. Do not apply excessive pressure with the
wipe, as this may release liquid that could damage the device internally. Do not insert the
wipe into the charging hole or sd slot. - Spread the Clinell wipe over the GSR and BVP sensors, always spreading the wipe
slightly with the sensors face down to ensure that no liquid leaks inside the device.
Versatile Bio (Run time: 5 minutes. Drying time: 1 minute).
To disinfect the sensors (GSR, Temperature, ExG...) use a Clinell wipe.
- Apply the wipe to the sensor head. Surfaces must remain wet for 1 minute counting the
application time. This application should be done lightly, always spreading the wipe with
the sensors upside down to ensure that no liquid leaks inside the device. - Apply the Clinell wipe to the cables, running from the sensor to the connector (do not
apply to the inside of the connector, as this could damage the electronics).
To disinfect the effort band sensor:
- For the blue fastening strap, immerse in 2% diluted Bomix. Be careful not to immerse the
sensor itself, only the fabric band.
To disinfect the amplifier:
- Apply the Clinell wipe to the entire housing of the device. Do not apply excessive
pressure with the wipe, this may release liquid which could damage the device internally.
Do not insert the wipe into the connector, charging hole, sd slot or phot/dig slot.
Storage
Keep the Ring on its original packaging when not in use.
If the device will not be used for an extended period, we recommend discharging the battery to 40-50% capacity. Storing the device for more than a week at 100% charge could cause permanent capacity loss.
Recommended battery storage temperature range:
-10ºC to 35ºC (14ºF to 95ºF) for up to 1 month
-10ºC to 30ºC (14ºF to 86ºF) for up to 6 months
Keep the Versatile Bio and every sensor on its original packaging when not in use.
If the device will not be used for an extended period, we recommend discharging the battery to 40-50% capacity. Storing the device for more than a week at 100% charge could cause permanent capacity loss.
Recommended battery storage temperature range:
-10ºC to 35ºC (14ºF to 95ºF) for up to 1 month
-10ºC to 30ºC (14ºF to 86ºF) for up to 6 months
Safety data sheets